MODEL 6711 BRAKİTERAPİ KAYNAĞININ MONTE CARLO SİMÜLASYONU KARAKTERİSTİKLERİ: ENERJİ SPEKTRUMU VE FANTOM BÜYÜKLÜĞÜNÜN ETKİLERİ

Berkay CAMGOZ 1 , Mehmet N. KUMRU 2

Brakiterapide kullanılan radyoaktif kaynakların, gerek doku içerisinde gerekse fantom metaryelindeki özelleşmiş doz dağılımlarının hassas olarak bilinmesi tedavi planlamalarında önemlidir. Prostat kanseri tedavisinde sıklıkla tercih edilen Model 6711 kaynağının dozimetrik özellikleri ve ilgili doz dağılımlarını ifade eden fonksiyonları literatürde verilmiş durumdadır. Bu çalışmada, bu kaynağa özel dozimetrik parametreleri etkileyeceği öngörülen radyoaktif malzemenin bozunum spektrumu ve kullanılan simülasyon fantomunun büyüklüğüne ait değişimler incelenerek etkileri Monte Carlo simulasyonu kullanılarak ortaya konmuştur. Bozunum spektrumu ve tek enerji kullanılarak yapılan çalışmalarda parametreler arasında %1 ile %68 oranında değişen farklar görülmüştür. Fantom genişliklerinin değişiminden kaynaklanan hesaplarda parametrelerin bire normalizasyonlarında beklendiği gibi %0,1’den büyük farklar gözlenmemiştir.

Keywords
Brakiterapi, Monte Carlo, Model 6711, EGSnrc, I-125,

MONTE CARLO SYMULATION CHARACTERISTICS OF MODEL 6711 BRACHYTHERAPY SEED: EFFECTS OF ENERGY SPECTRUM AND PHANTOM DIMENSIONS

Berkay CAMGOZ 1 , Mehmet N. KUMRU 1

It is important to determine specified dose distributions of radioactive seed sources of brachytherapy in both phantom material and treated tissues. Dosimetric features and specific dose distribution functions of Model 6711 brachytherapy seed which is frequently used in prostate cancer treatments, have been given in literature. In the present study decay spectrum of radioactive material and variations of phantom dimensions which are foreseen to affect dosimetric specifications of the seed, were investigated and results were performed by Monte Carlo simulation. In simulation using separate decay spectrum and mono energy, differences were noted between interval of 1-68 percent. Also in calculations of phantom dimensions variations, deviations were not observed more than 0.1 percent by normalization to 1 as expected.

Keywords
Brachytherapy, Monte Carlo, Model 6711, EGSnrc, I-125,

Details
   

Authors

Berkay CAMGOZ (1) (Corresponding Author)

Erciyes Üniversitesi
berkay.camgoz@ege.edu.tr

Mehmet N. KUMRU (2)

Ege Üniversitesi
lachry@yahoo.com

Supporting Institution

:

Project Number

:

Thanks

:
References